Description

The Additel ADT878-160-NO-110V is a reference dry-well temperature calibrator from the Additel 878 series with a control range of -40 to 160 deg C. Additel specifies a display accuracy of +/-0.1 deg C and a 30-minute stability of +/-0.005 deg C at full range. It suits calibration labs that need a reference-level dry well for temperature sensors and thermometers.

Specification Value
Series Additel 878 reference dry well calibrator, model ADT878-160
Temperature range -40 to 160 deg C (at 23 deg C)
Display accuracy +/-0.1 deg C at full range
Stability (30 min) +/-0.005 deg C at full range
Axial uniformity (60 mm) +/-0.035 deg C at -40 deg C; +/-0.020 deg C at 0 deg C; +/-0.050 deg C at 160 deg C
Axial uniformity (80 mm) +/-0.050 deg C at -40 deg C; +/-0.040 deg C at 0 deg C; +/-0.050 deg C at 160 deg C
Radial uniformity +/-0.01 deg C at full range
Loading effect +/-0.08 deg C (display sensor); +/-0.010 deg C (external sensor)
Communication Wi-Fi and Bluetooth; the data sheet also lists USB A, USB B and RJ45
Triple point of water Optional TPW kit for built-in automatic realization (ADT878-160 only)
Power 90-254 VAC, 45-65 Hz, 580 W

Source: Additel 878 reference dry well calibrator data sheet and product page (additel.com), checked 2026-10-28.